Customization: | Available |
---|---|
Standard or Nonstandard: | Standard |
Feature: | Corrosion-Resistant |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
disposable tips specification/ul | Dispense volume/ul | Dispense accuracy | Dispense CV |
50 (with filter element / cartridge filter) | 5 | 5.0% | 3.0% |
50 | 10 | 3.0% | 2.0% |
200 (with filter element / cartridge filter) | 10 | 6.0% | 3.0% |
200 | 50 | 2.0% | 0.75% |
1000 | 10 | 10.0% | 5.0% |
1. The test environment is 21-25ºC static wind environment; 2. The test mode is non-contact dispense method; 3. Replace the disposable tips with a new one after each full stroke. |
disposable tips specification/ul | Dispense volume/ul | Dispense accuracy | Dispense CV |
1000 (with filter element / cartridge filter) | 20 | 7.5% | 3.0% |
1000 | 50 | 5.0% | 2.0% |
200 | 10 | 5.0% | 3.0% |
1. The test environment is 21-25ºC static wind environment; 2. The test mode is non-contact dispense method; 3. Replace the disposable tips with a new one after each full stroke. |